Context
Within the field of MONALISA European LIFE project developed to validate a new method for pollen and allergen detection, the innovative Coriolis® continuous cyclonic air sampler is compared with the usual Hirst pollen trap
Protocol
•HIRST sampling
● Cylinder rotation 2 mm/h
● Air flow rate 10 L/min => Adhesive band
● Band recovered and cut out in 24h segments
● Segments coloured for microscopy
● Sample totality read by optical microscope x400
•CORIOLIS® sampling
●Air flow rate 200 l/min
●Sampling time 60 min => Liquid sample
●Centrifugation and elimination of the supernatant to keep a 2 ml residue
●Homogenization of the residue and preparation of 3 slides
●7 horizontal lines of the 3 slides read by optical microscope